Calibration Key for Suretemp 690/692 ThermometersCalibration Key for Suretemp 690 692 Thermometers Whenever the clinical accuracy of the thermometer is in question, verify the instrument calibration. It is recommended the unit be tested annually. This method uses a fixed value to verify that the device is calibrated properly. 1. Remove any probe currently connected to the SureTemp Plus Thermometer. 2. Connect the SureTemp Calibration Key. 3. Wait one minute for the device to enter Monitor mode and
Calibration Key for Suretemp 690/692 Thermometers
Whenever the clinical accuracy of the thermometer is in question, verify the instrument calibration.
It is recommended the unit be tested annually.
This method uses a fixed value to verify that the device is calibrated properly.
1. Remove any probe currently connected to the SureTemp® Plus Thermometer.
2. Connect the SureTemp® Calibration Key.
3. Wait one minute for the device to enter Monitor mode and display the temperature.
4. The temperature should read 36.3 +/- 0.1ºC or 97.3 +/- 0.2ºF.
5. Remove the SureTemp® Calibration Key.
6. Connect the original probe back to the device.
Shipping Notes
- Free Standard Shipping on $100+ Orders to the USA.
- Except Preorder products are shipped in 48 hours.
- Delivery to the USA:
- Standard Shipping : 3-10 business days
- If time is of the essence, please consider selecting expedited delivery for faster service.
Exchange/Return Notes
- We offer a 30-day return/exchange service after receiving.
- Final sale items are not eligible for returns or exchanges.
- To process your return/exchange, please contact us at [email protected]
- Please click here for more details>>> Return & Exchange Policy